Start

OUR PRODUCTS

Resuscitators
Cardiopulmonary 100A
Cardiopulmonary 100 


Resuscitatión Units
Case  90/6

Compressors
Compressor COMP

Oxygenoterapy Units
CPAP-AUTOCPAP-92/1
O2 Concentrator 2003
O2Mixer 99/1
O2 Economizer  ECO


Aspirators
Surgical  87/1  
Liposuction  87/1LP  

Surgical  99/4 
Continuous  89/15   
Obstetric  89/15 OB
Gastric Continuous 98/2
Gastric Intermittent 98/1
Pleural Neonatal  98/4
Portable 87/5/6/7 
Pedal aspirator
Obstetric Cup


Tourniquets
Electropneumatic 700A
Pneumatic 701   
Automatic 350  
Presure Distributions Valve
Sleeves


Venous Pump
Venous Pump 93/12

Continuous Flow Heater
Heater  91/11

Cranes
Hidraulic 95/3
Electric 95/4


Oxigen Plants
Oxigen Generator 2m3/h
Oxigen Generator 4m3/h
Oxigen Generating Plants
Machine for. Generating and Bottling oxygen
Oxigen Compressor OXICOMP

COMPRESSOR FOR OXYGEN
MICRO-CONTROLLED
OXICOMP-  FROM 4 TO 10 m3/h

Compressor for Oxygen

With the aim of solving the numerous problems currently arising when compressing oxygen and the frequent breakdowns of existing oxygen compressors, we in Ceimsa have applied our wide experience in this field to develop an oxygen compressor of controlled volume and low revolutions per minute (RPM), with a cooling system in the compression heads and in the piston liners so that the oxygen exits at low temperature thus lengthening the useful life of the machine and improving its performance.  It is a Micro-controlled Compressor of variable speed from 0 to18 m3/h designed to compress oxygen in two stages at 100 – 150 or 200 bars, programmable.

 

Its system of compressing oxygen with specially designed pistons, external stainless steel valves, liners and seals designed for heavy-duty work and manufactured with anti-explosive treatment make our compressor a unique machine in terms of its system, design, quality and sturdiness.

 

Our manufacturing processes include the latest technology to ensure oxygen compressors of the best quality and technological design making our compressors stand out from those of other manufacturers.

TECHNICAL CHARACTERISTICS

TECHNICAL CHARACTERISTICS OF THE TWO-STAGE OXYGEN COMPRESSOR  OXICOMP STANDARD

FLOW OF OXYGEN SUPPLY

0 to 18  m3/h with power supply pressure at 4 Bar

RATED VOLUME OF COMPRESSION OF THE PISTON

FIRST STAGE WITH PRESSURE

60 PSI

4 Bar

4 Kg/cm2

Piston volume 2.5 l x 4 Bars = 10 l of Oxygen Volume x 30 RPM = 300LPM x 60min =18m3/h

RATED VOLUME OF COMPRESSION OF THE PISTON

SECOND STAGE WITH PRESSURE

435 PSI

30 Bar

30 Kg/cm2

Piston volume 0.34 l x 30 Bars = 10 l of Oxygen Volume x 30 RPM = 300LPM X 60 min. = 18m3/h

BY MODIFYING THE POWER SUPPLY PRESSURE IT IS POSSIBLE TO OBTAIN OTHER RATED FLOWS OF SUPPLY, SUCH AS, FOR EXAMPLE WITH PRESSURE.

 

Using this formula we give warning of the increase or decrease in the supply flow insofar as the entry pressure should vary.

 

72 PSI

5 Bar

5 Kg/cm2

FIRST STAGE:

Piston Volume 2.5 l x 5 Bar = 12.5 l of Oxygen Volume x 30 RPM = 375 LPM x 60min = 22m3/h

536 PSI

37 Bar

37 Kg/cm2

SECOND STAGE:

Piston Volume 0.34 l x 37 Bar = 12,5 l of Oxygen Volume x 30 RPM = 375 LPM X 60 min. = 22 m3/h

MAXIMUM PRESSURE OF LOAD

2.900 PSI

200 Bar

204 Kg/cm2

RPM

30 RPM

Varying the RPM and the Kw. we can obtain other flows

OXYGEN CIRCUIT

Cooled by cooling liquid and air

POWER SUPPLY VOLTAGE

220–380V  50-60Hz

MOTOR POWER

10 Hp (7,5 Kw.)

WEIGHT

970 Kg.

DIMENSIONS

L 1.200 x B 600 x H 1.800 cm.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

POWER PRESSURE: (From 0 to 6 Bar)

This machine is supplied to work with an oxygen entry feed of 4 bars (this pressure can vary by more or less).  Once the pressure at which we want to work is set the compressor will stop if the pressure goes below the established pressure, giving warning that the pressure is insufficient while waiting for the pressure to rise.  When the pressure rises it will automatically become operative again.  Minimum and maximum pressures can be modified.

 

PRESSURE OF LOAD:  (100 - 150 or 200 Bar)

Setting the compressor in operation is designed so that the operator cannot commit any errors when selecting the filling pressure, for this reason, when switching on the machine a message appears on the display indicating that you must choose the desired loading pressure.  Having chosen the pressure another message appears to confirm the pressure chosen.  After confirming this, yet another message appears asking whether you wish to start the machine at the chosen pressure which means the operator must reconfirm before the compressor will start working.

 

CONTROLS:

These compressors are fitted with all kinds of electric and electronic controls of both pressure and temperature.  The temperature controls stop the compressor if there is overheating due to a lack of cooling liquid, excess heat or another cause.  Once the temperature has been established the machine will continue its cycle.

 

The Compressor will stop automatically and go on stand-by if there is no consumption, when the cylinders are full or there are no cylinders to be filled, and when the machine has reached the maximum pre-set pressure.

 

INSTRUCTIONS OF USE

The OXICOMP is normally used to fill oxygen cylinders but it can also fill cylinders and at the same time supply a Hospital Network at low pressure by means of a regulator.

 

 

MAINTENANCE: Our compressors, unlike those existing in the international market, have low cost maintenance, approximately 10% of their price over five years.

 

Every 5.000 hours operation you are recommended to clean the Non-return valves of the Compressor.  These valves can be checked by loosening the adaptors.  If any of them are not in good condition they must be replaced (the price of the valve is very economical).

 

Every 6.000 hours operation you must loosen and clean the pressure translators.

 

Very 10.000 hours operation you must check the  complete Pressure Modules.  For this reason you are advised to have a spare module for replacement in case of check-ups or possible breakdowns.

          

Example installation